Neuromuscular monitoring is used to evaluate neuromuscular function intraoperatively and to ensure complete neuromuscular recovery at the end of anaesthesia. Therefore, the lack of reliable neuromuscular monitoring devices that are not cumbersome to use is a major shortcoming for anaesthesia. A recently developed mechanomyography (MMG) device may meet these partially unmet needs due to its measurement of the patient's contractile force instead of its surrogates (i.e., acceleration, velocity), including the response to physiologically relevant tetanic stimulation. However, it is unclear whether the reliability of the newly developed MMG device is similar to or better than the currently available gold standard of neuromuscular monitoring based on electromyography (EMG).
The MEMORY trial is a randomized clinical agreement study which will prospectively enrol thirty ASA 1 or 2 patients ≥ 18 years scheduled for non-cardiac surgery requiring general anaesthesia and neuromuscular blockade for optimisation of surgical conditions. The patient's neuromuscular function is measured on one hand with MMG and the other hand with EMG in a randomised fashion. Additionally, randomisation will be stratified the domination of one hand. In the MEMORY trial we will test the hypothesis, that a newly developed MMG sensor provides not less reliable measurements of neuromuscular function compared to EMG in adult patients. Reliability includes agreement, precision, and the absence of staircase phenomenon and idiosyncratic responses to train-of-four (TOF) stimulation. In addition, we will explore if measurement of tetanic fade will provide more precise information about complete neuromuscular recovery than the current standard of TOF ratio \>0.9.
Study Type
OBSERVATIONAL
Enrollment
40
Precision of the Mechanomyography
Measured by the repeatability coefficient at baseline and at complete neuromuscular recovery
Time frame: intraoperative
Agreement of the TOF ratio between Electromyography and Mechanomyography
Agreement of the TOF ratio at 7 predefined TOF ratios: 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, and 0.9
Time frame: intraoperative
Confirmation of handling and precision of Mechanomyography sensor
Investigation of baseline stability (i.e., no staircase phenomenon of the first twitch response and the TOF ratio), occurrence and magnitude of idiosyncratic TOF ratio \> 1.0 during baseline measurement, repeatability of the TOF ratio during recovery from a rocuronium induced neuromuscular blockade, and sensitivity of the TOF ratio to changing the hand position during baseline measurements.
Time frame: intraoperative