The goal of this exploratory clinical trial is to evaluate the initial clinical safety and performance of a prototype wearable chest wall vibration (CWV) device intended to relieve exertional dyspnea in adults with Chronic Obstructive Pulmonary Disease. The main questions it aims to answer are: * Can CWV be delivered safely to participants via a wearable device? * Is there evidence of an effect on participant-reported dyspnea, endurance time and other measures of cardiorespiratory function? Participants will undergo two cycle-ergometer exercise testing sessions while wearing the prototype wearable device. The device will be active in one session (intervention) and inactive in the other (control). The order in which intervention or control occurs will be randomised. Researchers gather data relating to adverse events, device deficiencies, participant-reported symptom severity, endurance time and measures of cardiorespiratory function recorded via standard CPET apparatus.
The purpose of this Clinical Investigation is to evaluate the device design in a small number of human subjects with respect to initial clinical safety, and to gather preliminary data regarding the effect of the device on patient reported dyspnoea and physiological measures of cardiorespiratory function. This will facilitate planning of further steps of device development and validation, guide design modifications and define parameters for a future pivotal clinical investigation. The investigational device is at Pilot stage of Clinical Development, and the proposed Clinical investigation is a first-in-human exploratory study. It is not a superiority, non-inferiority, or equivalence study. The primary objective of this Clinical Investigation is to evaluate the investigational device with respect to initial clinical safety when used by subjects with Chronic Obstructive Pulmonary Disease during physical exertion in the form of cycle ergometer-based Cardiopulmonary Exercise Testing. The secondary objective of this Clinical Investigation is to gather preliminary data regarding the effect of the device on patient-reported dyspnoea and physiological measures of cardiorespiratory function. Each subject will attend the investigation site for a Pre-Study visit during which formal written consent will be obtained, clinical history and medications will be reviewed as part of standard CPET pre-assessment, and enrolment in the Clinical Investigation will be complete. An initial Incremental Cardiopulmonary Exercise Testing (CPET) session will then be conducted to determine baseline Peak Work Rate capacity. This is necessary so that the intensity of CPET that will be required to achieve 75% of each subject's individual Peak Work Rate capacity can be determined. Peak Work Rate capacity will be defined as the highest work rate that the Subject is able to maintain for ≥30 seconds. Following initial CPET, subjects will return for device testing on a non-consecutive day. Each subject will undergo 2 study arms during this visit: the Intervention (ResWave) and Control arms, in the random order. Both arms will be conducted on the same day, with a recovery period of 60 minutes between each arm. Subjects will undergo sub-maximal exercise testing at a Constant Work Rate (CWR) of 75% of their peak work rate capacity in both arms. Borg CR10 scores will be recorded at 2 minute intervals throughout. Endurance time will be recorded at test completion. Adverse Events will be recorded throughout testing and at 72-hour follow-up via phone call. Device Deficiencies will be recorded throughout testing.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
20
Prototype wearable in-phase chest wall vibration device
Beacon Hospital
Sandyford, Dublin 18, Ireland
Adverse Events
Adverse Events during or following use of the prototype device and/or exercise testing protocol
Time frame: 72 hours
Device Deficiencies
Performance data of the device during use
Time frame: During active use in exercise testing
Dyspnea severity
Borg Category-Ratio scale (Borg CR10). Scale from 0 (lowest intensity, better) to 10 (maximal intensity, worse).
Time frame: Recorded at 2 minute intervals during exercise testing. Outcome measure at isotime.
Endurance time
Exercise testing endurance time
Time frame: Duration of exercise testing
Reason for stopping
Whether participant stopped exercise testing due to dyspnea or other reason
Time frame: During exercise testing
Minute Ventilation (VE)
Minute Ventilation in millilitres/minute (mL/min)
Time frame: During exercise testing
Heart Rate (HR)
Heart rate in beats per minute (BPM)
Time frame: During exercise testing
Oxygen Pulse
Oxygen Pulse (VO2/HR) in millilitres/minute (mL/min)
Time frame: During exercise testing
Oxygen Consumption
Oxygen Consumption (VO2) in millilitres/minute (mL/min)
Time frame: During exercise testing
Carbon Dioxide Production
Carbon Dioxide Production (VCO2) in millilitres/minute (mL/min)
Time frame: During exercise testing
Work
Work (W) in watts
Time frame: During exercise testing
Tidal Volume (VT)
Tidal Volume in Litres (L) at body temperature and pressure saturated (BTPS)
Time frame: During exercise testing
Respiratory Quotient (RQ)
Respiratory Quotient (CO2 produced / O2 consumed; no unit as it is a ratio value)
Time frame: During exercise testing
End-tidal carbon dioxide tension (PETCO2)
End-tidal carbon dioxide tension (PETCO2) in millimetres of mercury (mmHg)
Time frame: During exercise testing
End-tidal oxygen tension (PETO2)
End-tidal oxygen tension (PETO2) in millimetres of mercury (mmHg)
Time frame: During exercise testing
Oxygen Saturation (SpO2)
Oxygen Saturation (SpO2) (%, percentage)
Time frame: During exercise testing
Respiratory Reserve
Respiratory Reserve in Litres
Time frame: During exercise testing
Expiration Time
Expiration Time in seconds
Time frame: During exercise testing