The study compares H2-BP, a wristband blood pressure monitor, with an auscultatory sphygmomanometer. Both devices are used on the same arm. After confirming wrist and upper arm circumference, H2-BP is operated by one measurer, while the auscultatory sphygmomanometer is simultaneously operated by two other measurers. Each subject undergoes alternating measurements of H2-BP on one wrist and the auscultatory sphygmomanometer on the same arm until three valid measurement pairs are obtained per subject, with a maximum of eight repetitions allowed. The two auscultatory sphygmomanometer measurers operate independently and cannot see each other's measurements during the test. Measurements are repeated for 10 to 30 minutes until three valid blood pressure values are obtained and recorded. Validity assessment variables are derived from the recorded values to evaluate the accuracy of H2-BP.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
141
The patient will have the test device applied to one wrist to measure blood pressure and the control device applied to the upper arm of the same arm to measure blood pressure. 1. H2-BP: The test device is placed on the wrist and presses a button to start pressurising. When the measurement is complete, the systolic and diastolic blood pressure is displayed on the screen. 2. big ben® Sphygmomanometer: The cuff is placed on the upper arm and the pump is pressed manually to inflate. 2 measurers will measure systolic and diastolic blood pressure by connecting 2 stethoscopes separately to one cuff. Due to the external nature of the study medical device, it is not possible to blind subjects and investigators, so that subjects and measurers are aware of the allocation. Therefore, blinding of subjects and measurers will not be used in this study.
Gangnam Severance Hospital,
Seoul, South Korea
Mean and Standard Deviation of Differences Between H2-BP Device and Aneroid Sphygmomanometer Measurements.
Average difference between blood pressure measurements obtained with the test medical device and the comparator. Standard deviation of the difference between blood pressure measurements obtained with the test medical device and the comparator (based on the number of measurements).
Time frame: serial assessment during 1 day
Standard Deviation of Differences Between H2-BP Device and Aneroid Sphygmomanometer Measurements.
Standard deviation of the difference in blood pressure measurements between the test device and the comparator (based on the number of participants)
Time frame: secondary outcome will be out in 1 day