This is a single-center, open-label, prospective study. Blood glucose is measured at different time points during oral glucose tolerance test and hypoglucose clamp test in healthy subjects, using both venous plasma and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS). Venous plasma glucose was set as gold standard. The two measurements were collected synchronously so as to calculate the mean absolute relative difference (MARD) and the consensus error grid (CEG). Accuracy of non-invasive blood glucose testing by mμSORS will be in-depth validated specifically in the 3-10 mmol/l glucose range.
In order to clinically verify the accuracy and safety of non-invasive blood glucose measurement by the mμSORS technology specifically in the 3-10 mmol/l glucose range, this study will conduct oral glucose tolerance tests and hypoglucose clamp tests on healthy subjects at different time points during the tests. In oral glucose tolerance tests, there will be 9 points as follows: 0-min and post glucose-load 30-min, 60-min, 90-min, 120-min, 150-min, 180-min, 210-min and 240-min. While in hypoglucose clamp tests, blood glucose will be set at around 3.0-4.0mmol/l for about 1.5-2.0 hours, blood glucose will be detected every 10 minutes.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
30
Blood glucose of participants were measured by venous plasma and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS). The two measurements were collected synchronously and analyzed.
Ruijin Hospital
Shanghai, Shanghai Municipality, China
Mean absolute relative deviation (MARD) of venous plasma glucose and glucose values measured by mμSORS at each time point of OGTT and hypoglucose clamp tests.
Glucose will be measured using both intravenous sampling (plasma) and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS) detection synchronously.
Time frame: 2 months
Consensus Error Grid (CEG) for venous plasma glucose and glucose values measured by mμSORS at each time point of OGTT and hypoglucose clamp tests.
Glucose will be measured using both intravenous sampling (plasma) and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS) detection synchronously.
Time frame: 2 months
MARD for two measurement methods in different blood glucose ranges.
Plasma glucose ranges 3.9-10.0 mmol/L and \<3.9 mmol/L.
Time frame: 2 months
Incidence of Treatment-Emergent Adverse Events.
Safety outcomes.