Perimetry is an essential component in the diagnosis and monitoring of glaucoma. Since the advent of Standard Automated Perimetry (SAP), one of the clinical standards has been the Humphrey Field Analyzer (HFA, Carl Zeiss AG, Oberkochen, Germany). Visual field (VF) testing provided by the HFA is standard-of-care in glaucoma clinics and other ophthalmology and optometry practices. However, the HFA is a large device that does not allow for examination outside the clinic and can be uncomfortable for patients with limited mobility. This study aims to evaluate a novel head-mounted perimeter against the HFA as an alternative method of VF testing. This new perimeter uses a Virtual Reality (VR) headset and a gamified version of VF testing to assess the visual function of healthy eyes and patients. The results of this study will potentially serve as pilot data for the design of a larger study that involves the full assessment of the VR headset and its VF test, based on various testing strategies.
Procedures: 1. Ophthalmic screening: (a) Visual acuity, slit lamp exam, rebound tonometry (all subjects) 2. OCT examination of the study eye with the Topcon Maestro2 Optical Coherence Tomography (OCT): (a) Standard Retinal Nerve Fiber Layer (RNFL), Optic Disc and Ganglion Cell Complex (GCC) scan patterns (i) Fundus photo is automatically obtained. (b) Abnormalities consistent with ophthalmic disease will be an exclusion for the healthy group of subjects. Reasonable and common artifacts that do not significantly impair the scan report will be accepted. (c) OCT abnormalities consistent with glaucoma are expected on the 30 glaucoma subjects. 3. Once eligible, each patient will perform four (4) visual field tests: (a) once with the HFA Swedish Interactive Threshold Algorithm (SITA) Standard 24-2 pattern (i) Reliability criteria: False Positives no higher than 15% (b) three tests with the VR headset: (i) two tests with ZEST strategy (ii) one test with gamified ZEST strategy (c) Counter-balancing will be used to determine the order of (a) and (b). In particular, a study participant will either start with the HFA visual field test and then proceed to the three tests with the VR headset; or, start with the three VR tests and finish with the HFA visual field test. The sequence of the visual field tests for each study ID will be provided by the sponsor.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
60
This study aims to evaluate a novel head-mounted perimeter against the HFA as an alternative method of VF testing. This new perimeter uses a Virtual Reality (VR) headset and a gamified version of VF testing to assess the visual function of healthy eyes and patients.
Illinois College of Optometry / Illinois Eye Institute
Chicago, Illinois, United States
RECRUITINGVisual field perimetry light sensitivity
Localized differential light sensitivity threshold measured in decibels (dB)
Time frame: 1 day
Visual field perimetry foveal sensitivity
Foveal threshold in decibels (dB)
Time frame: 1 day
Visual field perimetry test duration
Test duration measured in minutes
Time frame: 1 day
Optical Coherence Tomography (OCT)
Optical Coherence Tomography (OCT) thickness values may be used, such as Retinal nerve fiber layer (RNFL) and ganglion cell complex (GCC) thickness, measured in microns.
Time frame: 1 day