The REQUEST study aims to evaluate the use of a new wireless monitoring device, called the viQtor solution, on a surgical ward in Catharina Hospital Eindhoven. This device continuously tracks vital signs, such as heart rate, breathing rate, and blood oxygen levels, in patients after surgery. The goal is to see if the viQtor solution can help detect health problems earlier and reduce the workload for nurses by minimizing the need for manual checks. The study will involve 500 patients and will take place over 7 months.
Detailed Description This is a prospective implementation evaluation study of the viQtor solution, a wearable medical device designed for continuous monitoring of vital signs in post-operative patients. The study will be conducted at the Catharina Hospital Eindhoven on the surgical ward "5 Oost" and will be divided into two phases, each involving approximately 250 patients. In Phase 1 (Months 1-3), patients will continue to receive standard care, which includes manual spot checks and Modified Early Warning Score (MEWS) assessments by nursing staff. During this phase, healthcare professionals will receive training on how to use the viQtor solution. In the Evaluation and Validation Period (Month 4), the accuracy of the viQtor device will be validated by comparing its continuous monitoring data with the manual measurements taken by nurses. In Phase 2 (Months 5-7), the viQtor solution will become the primary method for monitoring patients, with manual checks conducted only when necessary. The study aims to assess the viQtor solution's usability, acceptance, adoption by healthcare professionals, and its accuracy in detecting patient deterioration. It will also evaluate the feasibility of implementing this technology in a real-world hospital setting and its impact on clinical outcomes. The viQtor solution is a CE-marked Class IIa medical device, ensuring compliance with European safety standards. It is designed to be worn on the upper arm and is capable of wirelessly transmitting data every five minutes to a cloud-based platform accessible to healthcare professionals. Participants in this study will include adult patients admitted to the surgical ward for various procedures. The study will not only assess the technical performance of the viQtor solution but also its impact on the workflow and workload of healthcare providers. Data collected during the study will be used to determine the potential benefits and challenges of implementing continuous monitoring technology in hospital settings. The ultimate goal is to improve patient safety and outcomes through early detection of health deterioration, while also reducing the burden on nursing staff. If successful, this study could pave the way for broader adoption of wearable monitoring technology in hospitals, potentially setting a new standard of care for post-operative patients. This detailed description provides a comprehensive overview of the study objectives, methodology, and expected outcomes without duplicating information entered elsewhere in the record.
Study Type
OBSERVATIONAL
Enrollment
500
The viQtor device is a wearable medical device designed for continuous monitoring of vital signs, including heart rate (PR), respiratory rate (RR), and blood oxygen saturation (SpO2). It is worn on the upper arm and transmits data wirelessly to a cloud-based platform. This device is being evaluated as part of the standard monitoring protocol for post-operative patients in a surgical ward setting.
Catharina Hospital Eindhoven
Eindhoven, North Brabant, Netherlands
Usability of the viQtor Monitoring Solution
Evaluation of the usability of the viQtor solution based on the System Usability Scale (SUS). The outcome will be measured by the SUS score, with a target score of \>68 indicating acceptable usability.
Time frame: 8 months
Reduction in Nursing Workload
Assessment of the reduction in nursing workload by comparing the time spent on manual spot checks before and after implementing the viQtor solution. Success is defined as a 20% reduction in time spent on manual checks.
Time frame: 8 months
Predictive Accuracy of the viQtor Solution
Evaluation of the viQtor solution's accuracy in predicting patient deterioration by comparing the continuous vital sign measurements (SpO2, PR, RR) with intermittent measurements conducted as part of the MEWS protocol.
Time frame: 8 months
Implementation Success of the viQtor Solution
Determination of the successful implementation of the viQtor solution, measured by the percentage of available and usable data (\>70%) collected from the device during the study period.
Time frame: 8 months