This study shall be a prospective, multicenter, single arm, consecutive, interventional study conducted in a post-market setting using commercially available devices. Consecutive, eligible patients with clinical suspicion of MASLD or MASH reporting for an endoscopic ultrasound and liver biopsy for evaluation of fibrosis will be enrolled. EUS Shear Wave Elastography and Attenuation Imaging technologies will be compared to liver biopsy and FibroScan results and other non-invasive fibrosis screening modalities . The data collected during this study will be evaluated in accordance with the procedures set forth in the protocol. The main question\[s\] it aims to answer are: * Establish optimal cutoffs for EUS-SWE in reference to liver biopsies staging system for liver fibrosis * Evaluate the diagnostic performance of EUS-SWE compared to FibroScan (VCTE) and to other non-invasive fibrosis screening modalities (screening scores). Participants will undergo: * Endoscopic Ultrasound with Shear Wave Elastography (SWE) and Attenuation Imaging (ATI) * Liver biopsy * FibroScan
A total of up to 300 subjects will be enrolled and treated at up to six US sites. Enrollment is expected to be completed within 18 months. Each subject will undergo procedures (EUS with SW Elastography and Attenuation Imaging, liver biopsy and FibroScan). Liver biopsy and FibroScan should be done within three months from Endoscopic ultrasound with SWE and ATI. End of study participation for a subject will be reached after pathology results have been obtained. This study has been designed as a post-market clinical registry to evaluate the effectiveness of Endoscopic Ultrasound with Shear Wave Elastography to estimate hepatic fibrosis in patients with chronic liver disease. All consenting, eligible patients reporting for an ultrasound and liver biopsy for evaluation of fibrosis meeting the inclusion criteria will be consecutively enrolled into the study and assigned a unique study identification number.
Study Type
OBSERVATIONAL
Enrollment
300
EUS combines ultrasound technology with endoscopy to gain an internal vantage point to obtain higher resolution ultrasound images. Shear Wave is a software option on the ultrasound system EUS that provides an advanced measure and dynamic display of tissue stiffness which is being evaluated for use in assessing liver fibrosis. Attenuation Imaging (ATI) is a novel application that provides the capability to quantify and color-code the changes in attenuation coefficient of the liver that may arise with changes in liver composition (e.g., increased fat levels).
A small ultrasound probe to measure the velocity of a sound wave passing through the liver. This is converted into kilopascals (kPa), which is used to assess liver fibrosis.
A biopsy of liver tissue could be obtained under endoscopic ultrasound guidance, interventional radiology (percutaneously), or surgically. Endoscopic Ultrasound-guided liver biopsy provides high resolution images of left lobe of the liver and a good portion of the right lobe of the liver. This, coupled with Doppler capability, allows the physician to direct the biopsy needle safely into the liver for sampling under real time image guidance.
Brigham and Women's Hospital
Boston, Massachusetts, United States
RECRUITINGGeisinger Medical Center
Danville, Pennsylvania, United States
RECRUITINGShear Wave Elastography (SWE) vs liver biopsy METAVIR staging system
Establish optimal cutoffs for EUS-SWE cutoffs in reference to liver biopsies on METAVIR staging system for significant fibrosis (F0-1 vs F2-4), advanced fibrosis (F0-2 vs F3-4), and cirrhosis (F0-3 vs F4) based on area under the receiver operator characteristics (AUROC) curve analysis.
Time frame: During study data analysis
SWE vs FibroScan and other non-invasive fibrosis screening modalities ( screening scores)
Evaluate the diagnostic performance of EUS-SWE compared to FibroScan (VCTE) and to other non-invasive fibrosis screening modalities (e.g., Fibrosis-4 Score (FIB-4), AST to Platelet Ratio Index (APRI), and NAFLD Fibrosis Score (NFS)).
Time frame: During study analysis
Correlation of EUS-SWE to fibrosis staging (from liver biopsy).
Evaluate the correlation of EUS-SWE to fibrosis staging (from liver biopsy).
Time frame: During study analysis
Correlation of EUS-ATI to steatosis score (from liver biopsy)
Evaluate the correlation of EUS-ATI to steatosis score (from liver biopsy).
Time frame: During study analysis