The study will be conducted in hospitalised patients who undergo spot-check monitoring of vital signs as part of their routine care.
In an additional spot-check measurement procedure, the study will measure oxygen saturations through different Philips SpO2 finger sensors. Patients will wear a nasal cannula linked to a CO2 measuring device at the same time. The data from both devices will be used to design and improve algorithms for measurement of respiratory rate with the saturation probe and the Philips FAST pulse oximeter system. The study will include 50 adult patients and 25 pediatric patients.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
86
Spot-check measurement: different Philips SpO2 sensors and Philips FAST pulse oximeter system, capnography
Ysbyty Gwynedd Hospital
Wales, Bangor, United Kingdom
Data collection for database for algorithm development
Anonymized database of subject data from the Oxygen Saturation finger sensors (photoplethysmogram waveform) and the reference sensors (capnography waveform)
Time frame: 6 months