The purpose of the research study is to investigate the peak sacral interface pressures of two separate hospital bed systems in the supine and 30 degrees head-of-bed elevation position.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
18
Low air loss mattress in supine position
Standard hospital mattress in supine position
Low air loss mattress with head of bed at 30 degrees
University of Alabama
Birmingham, Alabama, United States
Peak Sacral Pressure
Peak sacral pressure is defined as the highest pressure within a 9-10 cm squared area under the coccyx. The area under coccyx was identified prior to each condition tested using the pressure mapping system. Presure was recorded continuously for 15 minutes. Because the mattresses tested were advertised as 'cycling' pressure, with the cycles set to 5 minutes, the average of minutes 1-3 and 6-8 were used for analysis. These periods represent different phases of the 'cycling' where sacral pressure would conceivably differ.
Time frame: minutes 1-3 and minutes 6-8
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Standard hospital mattress with head of bed at 30 degrees