The DiaSorin Molecular LIAISON® NES FLU A/B, RSV \& COVID-19 real-time polymerase chain reaction (RT-PCR) assay is intended for use on the DiaSorin LIAISON® NES instrument for the in-vitro qualitative detection and differentiation of nucleic acid from influenza A, influenza B, RSV and SARS-CoV-2 virus from dry nasal swabs (NS) from human patients with signs and symptoms during the acute phase of respiratory tract infection in conjunction with clinical and epidemiological risk factors. The LIAISON® NES FLU A/B, RSV \& COVID-19 assay is intended for use as an aid in the differential diagnosis of influenza A, influenza B, RSV and SARS-CoV-2 infection in a professional laboratory setting. Negative results do not preclude influenza A, influenza B, RSV or SARS-CoV-2, infection and should not be used as the sole basis for patient management decisions. The assay is not intended to detect the presence of the influenza C virus.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
1,000
The LIAISON® NES FLU A/B, RSV \& COVID-19 assay used on the LIAISON® NES instrument is a real-time PCR system that enables the direct amplification, detection and differentiation of Flu A viral RNA, Flu B viral RNA RSV viral RNA, and SARS-CoV-2 RNA from dry nasal swabs. Nasal Swabs can be professionally collected by a Healthcare Provider or self-collected by the patient under the supervision of the healthcare provider. The collected nasal swab can then be directly loaded into the cartridge without nucleic acid extraction.
PAS Research - Henderson
Henderson, Nevada, United States
RECRUITINGPAS Research - Pittsburgh
Pittsburgh, Pennsylvania, United States
RECRUITINGPAS Research - McAllen
McAllen, Texas, United States
RECRUITINGDiagnostic Accuracy
Estimates of Sensitivity/percent positive agreement (PPA) and Specificity/percent negative agreement (NPA) will be calculated based on a two-by-two table (comparator method result vs. result from LIAISON® NES FLU A/B, RSV \& COVID-19 assay or LIAISON PLEX® RSP Flex assay) for each target. In addition, 95% two-sided confidence intervals will be provided. A separate Sensitivity/PPA and Specificity/NPA for each target, to include those specimens excluded due to a discrepancy between the standard-of-care and molecular comparator result, will be presented in a separate 2 x 2 table.
Time frame: Samples will be tested on LIAISON NES within two hours of collection.
Clinical Performance
The secondary objective is to compare the clinical performance of the LIAISON NES FLU A/B, RSV \& COVID-19 assay per clinical site, target patient sub-population (e.g. age group, patient location) and specimen collection method (e.g. professionally collected, patient self-collected).
Time frame: Samples will be tested on LIAISON NES within one hour of collection.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.