Systematically evaluate the safety performance, intelligent application effect and experience of PPE developed in different scenarios.
In high-level biosafety laboratories, the National Center for Infectious Disease Medicine, medical consortiums and other national platforms, systematic assessment of the project developed individual physical protection equipment in different scenarios of the safety performance, the intelligent application of the effect and the use of experience, the assessment of the indicators covering the protection performance, safety, stability, comfort, environmental protection, intelligence, ease of operation, including but not limited to local breakage protection performance, Aerosol protection performance, airtightness, liquid penetration performance, differential pressure stability, noise value, air supply flow, ease of wear and removal, ease of movement, ease of operation, degradability, degree of intelligence, etc., and ultimately determine the technical indicators and testing methods of individual physical protective equipment, and establish evaluation technical standards. Based on the determined technical indexes and testing methods of individual physical protective equipment, we will carry out parallel comparative evaluation of similar foreign imported equipment, and form a technical research report on the evaluation of the performance of relevant protective equipment. Optimise the quantitative performance indexes of PPE for domestic shortcomings and weaknesses. Prepare standard operation norms and technical standards for individual physical protective equipment.
Study Type
OBSERVATIONAL
Enrollment
100
Capital medical university
Beijing, Beijing Municipality, China
Performance evaluation of medical protective clothing
Evaluate the performance of self-developed equipment versus commercially developed equipment, such as protective clothing, hoods, boots and gloves. Protective clothing: Surface moisture resistance: according to the standard including GB/T 4745-1997 on the level of water, the protective clothing outside the water level is not less than level 3 Hydrostatic pressure: according to GB/T 4744-1997, the hydrostatic pressure is not less than 1.67kPa or 17cmH2O. Moisture permeability: according to GB/T 12704-1991, the moisture permeability is not less than 2500g/(m²-24h). Water staining test: according to GB/T 4745-1997, the outer water staining level is not less than 3. Microbiological barrier performance: the protective clothing should meet the requirements of microbiological indicators in GB 15979-2002 to ensure that it can effectively block the invasion of bacteria, viruses and other microorganisms.
Time frame: 3 years
Protective headgear performance evaluation
Protective hood: Bacterial filtration efficacy expressed as BFE Rupture strength and elongation at break evaluated in Newtons Moisture permeability is evaluated by m²-24h. Water exposure test: According to GB/T 4745-1997, the external water exposure level should not be less than grade
Time frame: 3 years