The purpose of the study is to follow-up on a released low-risk device as part of standard PMCF activities
This is a comparative study which is conducted monocentric at Hoerzentrum Oldenburg, Germany. The focus is to test a released hearing aid and compare this to the subject's own hearing aids with a minimum of 16 mild-to-moderate sensorineural hearing impaired adult subjects. The subjects, all of them native German speakers, have 2 appointments of 1-2 hours including a fitting procedure with real-ear measurements and some standard speech tests and undergo two 2-weeks home trials (with questionnaires).
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
20
Receiver in Canal (RIC) hearing aid, M Receiver
subjects' own hearing aids, any brand, any model
Hörzentrum Oldenburg gGmbH
Oldenburg, Germany
Oldenburg Sentence Test
The Oldenburg Sentence Test (OLSA) is an audiometric test for determining the speech recognition threshold (SRT in dB). Sentences of the form name - verb - numeral - adjective - noun are used as speech material. The sequence of words is a random combination from an inventory of 50 words in total. The design of the test prevents memorization of the sentences, so the OLSA can be repeated as often as desired. For this study speech in noise recognition was measured. Noise level was 65 dB SPL played from the back, speech was presented from front. The SRT shows the signal-to-noise ratio needed for 80% speech recognition. Lower values mean better speech understanding (i.e. the speech can be quieter with respect to the noise and still be understood). There are no minimum or maximum values.
Time frame: Estimated Time Frame 2 weeks (Test conducted at second lab session).
Freiburger Monosyllabic Speech Test
To investigate speech intelligibility (SI) performance for speech at 65dB in a quiet listening situation, the Freiburger monosyllabic speech test is used which is still the gold standard for Hearing Care Professionals (HCPs) in Germany. Results are reported in Percent correct data. Minimum is 0%, maximum is 100% with higher values meaning better speech understanding.
Time frame: Estimated Time Frame 2 weeks (Test conducted at second lab session).
Match-to-Target
Investigate how close the study hearing aids fitted and fine-tuned with the fitting software to NAL-NL2 Targets match the postulated targets, as measured in-situ with real-ear measurements. Absolute deviations were calculated at specific frequencies ranging from 250 Hz to 4 kHz. Unit of measure is dB SPL. Minimum Value is 0 (i.e. measured value and target value identical), no maximum value. Higher values mean greater deviation from target.
Time frame: 1 day (Fitting procedure with real-ear measurements is done at first lab session)
Speech, Spatial and Qualities of Hearing Questionnaire (SSQ)
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
A short form of the SSQ is used (SSQ 17, German version), to investigate speech perception, sound localisation, quality of hearing perceived in real-life. Questions are answered on a scale from 0 to 10, higher scores mean better outcomes. The responses to the individual questions are grouped for the subsections (speech, spatial, quality) by calculating the means.
Time frame: 2 weeks
International Outcome Inventory for Hearing Aids (IOI-HA)
Investigate outcomes and effectiveness of hearing aid usage in real-life. Responses are given on a scale from 1 - 5. Here, the mean of questions 2 -7 is reported. Higher values represent better outcomes. Question 1 was on wearing time, which is not considered a rating of effectiveness and was therefore excluded from the mean and will be reported separately. Fpr wearing times higher values represent longer wearing time (maximum 5 equals "more than 8 hours a day").
Time frame: 2 weeks
Satisfaction Questionnaire
Investigate satisfaction in real-life for different aspects such as sound quality, naturalness of sound, conversation with one person and overall satisfaction. Each aspect was assessed on on a 7-point likert scale, going from 1 (very dissatisfied) to 7 (very satisfied) The overall satisfaction and sound quality overall were individual questions and are not the combination of different situations/aspects. These questions were also assessed on on a 7-point likert scale, going from 1 (very dissatisfied) to 7 (very satisfied). Higher values represent better outcomes.
Time frame: 2 weeks