You were selected to participate in this observational study because you suffer from glaucoma and cataracts. For the cataract you will have cataract surgery. This study will not affect your treatment, it will only collect data on what is performed before, during and after the procedure. The aim of this data collection is to demonstrate the safety of a new generation of cataract surgical devices in patients who also suffer from glaucoma. They are glaucoma patients who have an increased sensitivity to eye pressure increases both during and after surgery. The new surgical device makes it possible to safely perform cataract surgery under low pressures. This could, at least theoretically, provide an advantage for patients who also suffer from glaucoma.
Cataract surgery is a highly standardized procedure. However a new phacoemulsification system "Quatera 700" has been developed by Carl Zeiss Meditec. The system has a unique patented QUATTRO Pump that directly controls infusion and aspiration and enables a synchronous change in both. It has four membranes, which directly regulate infusion and aspiration and maintain a pre-set intraocular pressure (IOP), resulting in a stable anterior chamber during the surgery. The membranes are activated by pneumatic pressure regulated by sensors that measure infusion and aspiration flows which allows the system to determine and compensate for incisional leakage as well. This innovative pump system allows surgeons to significantly decrease the intra-ocular pressure during cataract surgery. In respect to competitive devices on the market, the Quatera system is able to operate with pressure four times lower than average. Especially in glaucoma patients (patients suffering from optic nerve head damage due to elevated intraocular pressures), performing surgery with lower intra-ocular pressures would help in avoiding extra nerve fiber layer damage and overall safety of the surgery.
Study Type
OBSERVATIONAL
Enrollment
10
Visionair Oogzorg
Geel, Belgium
RECRUITINGComplication rate during cataract surgery
Complication rate during cataract surgery
Time frame: Month 1
Best corrected visual acuity
Best corrected visual acuity will be determined and recorded at the preoperative, day 1, week 1 and month 1 visit.
Time frame: Month 1
Intraocular pressure
The intraocular pressure will be measured at the preoperative, day 1, week 1 and month 1 visit. Ocular hypertension will reported.
Time frame: 1 month
Retinal nerve fiber layer thickness
Optical Coherence Tomography will be performed at the preoperative and month 1 visit. Any abnormalities will be described.
Time frame: 1 year
Visual field mean deviation
The visual field examination will be performed at the preoperative and month 1 visit. Any abnormalities will be described.
Time frame: Month 1