The purpose of this research is to compare the performance of acceleromyograph and electromyograph-based quantitative neuromuscular monitors when the arms are tucked for surgical positioning.
Study Type
OBSERVATIONAL
Enrollment
40
Simultaneous readings from electromyograph and acceleromyograph
Mayo Clinic
Jacksonville, Florida, United States
Accuracy and precision
Accuracy and precision of train-of-four ratios measured simultaneously by two quantitative neuromuscular monitors every minute among 40 patients undergoing surgery of at least 30 min duration (roughly 1200 measurements) in which the arms are tucked and in the absence of neuromuscular blocking agents. In the absence of neuromusuclar blocking agents, the expected train-of-four ratio is 1.0. Accuracy of each monitor will be assessed by how close measuremetns are to 1.0 while precision will be assessed by the variability of the measured responses in all patients undergoing procedure.
Time frame: 120 minutes