Comparing a wireless accelerometer-based sensor that continuously measure breathing rate with the gold standard capnography. The aim is to evaluate if both methods are equivalent.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
PREVENTION
Masking
NONE
Enrollment
29
This sensor is being fasten on the chest of the patients and measures respiratory rate with the help of an accelerometer. Data is being sent wireless to a backend monitor.
Haraldsplass Deaconess Hospital
Bergen, Norway
Validation of correctness of the sensor
The respiratory rate measured by the sensor is being compared to the respiratory rate by capnography. A discrepancy of +- 3 breaths/min is within the acceptable range.
Time frame: one hour