The study will evaluate the measuring pulse rate (PR) using the Norbert Device (ND) as compared to a reference heart rate measured using an electrocardiogram (ECG).
The study design is to validate the pulse rate accuracy. The study is a non-randomized, non-blinded single arm design to evaluate the clinical accuracy of the ND's pulse rate measurement compared to a reference heart rate manually measured by observing R-peaks in a standard concurrently recorded ECG from an FDA 510(k)-cleared reference clinical ECG (RCECG). The sponsor will ensure that the study staff enrolls a range of subjects.
Study Type
OBSERVATIONAL
Enrollment
40
The Norbert Device (ND) is a contactless, noninvasive device that provides measurements of pulse rate.
Vigilant Clinical Testing
Irvine, California, United States
Primary Objective
To validate the accuracy specification for pulse rate from the ND when compared to an FDA-cleared medical device for heart rate monitoring electrocardiogram (ECG monitor, 5 lead). Acceptance criteria for the overall cohort is ≤ 3.0 bpm for PR in subjects compared to a gold standard ECG.