The aim of the study is to assess the accuracy of the automatic oscillometric BP measuring device at the brachial level, the WITHINGS BPM Pro 2, in the general population
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DEVICE_FEASIBILITY
Masking
NONE
Enrollment
89
The patient will be measured alternatively with a reference mercury sphygmomanometer (four times) and the device under study (3 times)
N2
Yerevan, Armenia
Blood pressure measurement performance evaluation per measure
The blood pressure (systolic and diastolic blood pressure) measurement performance will be established using the mean value of the differences and the standard deviation between the device under study and the reference for each measure.
Time frame: The time frame is three months.
Blood pressure measurement performance evaluation by subject
The blood pressure measurement performance by subject will be established using the mean value of the differences and the standard deviation between the averaged per subject value of blood pressure (systolic and diastolic blood pressure) device under study and its reference.
Time frame: The time frame is three months.
Safety use of the device
The rate of adverse effects.
Time frame: The time frame is three months.