The goal of this open label study is to evaluate the efficacy of smart phone-based test (SOBEREYE OPTOVERA) on measuring pupillary light reflex (PLR) alterations following cannabis use healthy in adults. The main question it aims to answer is: Can SOBEREYE OPTOVERA detect PLR alterations following cannabis consumption in healthy adults, in comparison to a pupillometer? Participants will be given two 5 mg capsules to be ingested for the 10 mg THC dose or five 5 mg capsules to be ingested for the 25 mg THC. Participants will be asked to complete PLR tests throughout the study day.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
20
10 mg of Tetrahydrocannabinol (THC) provided as a softgel capsule
25 mg of THC provided as a softgel capsule.
SOBEREYE OPTOVERA is a portable, non-invasive test that measures the Pupillary Light Reflex (PLR)
NeuroLight is an automated pupillometer that generates a flash of light and measures the photomotor reflex very accurately.
KGK Science Inc.
London, Ontario, Canada
RECRUITINGThe difference in change in Pupillary Light Reflex (PLR) measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=5 min.
Time frame: T=0 to T=5 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=30 min.
Time frame: T=0 to T=30 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=60 min.
Time frame: T=0 to T=60 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=120 min.
Time frame: T=0 to T=120 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=180 min.
Time frame: T=0 to T=180 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=240 min.
Time frame: T=0 to T=240 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=300 min.
Time frame: T=0 to T=300 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=360 min.
Time frame: T=0 to T=360 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=420 min.
Time frame: T=0 to T=420 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=480 min.
Time frame: T=0 to T=480 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=5 min.
Time frame: T=0 to T=5 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=30 min.
Time frame: T=0 to T=30 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=60 min.
Time frame: T=0 to T=60 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=120 min.
Time frame: T=0 to T=120 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=180 min.
Time frame: T=0 to T=180 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=240 min.
Time frame: T=0 to T=240 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=300 min.
Time frame: T=0 to T=300 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=360 min.
Time frame: T=0 to T=360 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil reflexes (absolute contraction and latency) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=480 min.
Time frame: T=0 to T=480 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer in pupil size from time (T)=0 minutes (min) at T=5 min.
Time frame: T=0 to T=5 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=30 min.
Time frame: T=0 to T=30 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=60 min.
Time frame: T=0 to T=60 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=120 min.
Time frame: T=0 to T=120 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=180 min.
Time frame: T=0 to T=180 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=240 min.
Time frame: T=0 to T=240 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=300 min.
Time frame: T=0 to T=300 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=360 min.
Time frame: T=0 to T=360 minutes
The difference in change in PLR measures between a smart phone-based test and NeuroLight pupillometer
The difference in change in pupil activity as measured by pupillary unrest in ambient light (hippus) between a smart phone-based test and NeuroLight pupillometer from time (T)=0 minutes (min) at T=480 min.
Time frame: T=0 to T=480 minutes
The difference in change in reaction time
The difference in change in reaction time as assessed by Dynavision from T= 0 at T=60 minutes
Time frame: T= 0 at T=60 minutes
The difference in change in reaction time
The difference in change in reaction time as assessed by Dynavision from T= 0 at T=120 min
Time frame: T= 0 at T=120 minutes
The difference in change in reaction time
The difference in change in reaction time as assessed by Dynavision from T= 0 at T=180 minutes
Time frame: T= 0 at T=180 minutes
The difference in change in reaction time
The difference in change in reaction time as assessed by Dynavision from T= 0 at T=240 minutes
Time frame: T= 0 at T=240 minutes
The difference in change in reaction time
The difference in change in reaction time as assessed by Dynavision from T= 0 at T=300 minutes
Time frame: T= 0 at T=300 minutes
The difference in change in reaction time
The difference in change in reaction time as assessed by Dynavision from T= 0 at T=360 min
Time frame: T= 0 at T=360 minutes
The difference in change in reaction time
The difference in change in reaction time as assessed by Dynavision from T= 0 at T=420 min
Time frame: T= 0 at T=420 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T= 5 min, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: T=0 at T= 5 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=30 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=30 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=60 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=60 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=90 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=90 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=120 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=120 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=180 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=180 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=240 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=240 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=300 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=300 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=360 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=360 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=420 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=420 minutes
The difference in change in subjective drug effects
The difference in change in subjective drug effects as assessed by the modified Drug Effects Questionnaire (DEQ) from T=0 at T=480 minutes, on a scale of 1 to 10. Higher score indicates higher subjective drug effect.
Time frame: T=0 at T=480 minutes