This is a prospective, randomized study to compare the effects of Bachmann bundle pacing (BBP) versus conventional right atrial appendage (RAA) pacing on atrial electromechanical synchrony in patients with sick sinus syndrome (SSS). The study utilizes two-dimensional speckle tracking imaging (2D-STI) to assess improvements in cardiac function and synchrony, aiming to determine a more optimal pacing site for SSS patients.
Sick sinus syndrome (SSS) is a common indication for permanent pacemaker implantation. Conventional right atrial appendage (RAA) pacing, however, is non-physiological and may lead to atrial conduction delays, atrial remodeling, and an increased risk of atrial fibrillation. Bachmann bundle pacing (BBP) is emerging as a more physiological pacing strategy that may correct atrial conduction delays and improve electromechanical synchronization. This study prospectively enrolled and randomized 72 patients with SSS to receive either BBP or RAA pacing. The primary objective is to evaluate the differences in atrial synchrony, cardiac function, and pacing parameters between the two groups at baseline, 1-month, and 3-month follow-ups, using electrocardiography and two-dimensional speckle tracking imaging (2D-STI). The study aims to provide evidence for BBP as a superior treatment option for improving long-term outcomes in SSS patients.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
72
Implantation of a permanent pacemaker system. The atrial lead (Medtronic 3830 active electrode) is surgically positioned and fixed in the Bachmann bundle region, located at the angle between the superior vena cava and the atrium in the mid-posterior position of the high atrial septum. Successful placement is confirmed by achieving a P-wave width shortening of at least 30ms compared to the preoperative P-wave. The ventricular lead is positioned in the mid-to-lower portion of the ventricular septum. Pacing parameters are measured and confirmed to be good.
Implantation of a permanent pacemaker system using the conventional technique. The atrial active electrode is surgically positioned and fixed in the right atrial appendage. The ventricular lead is placed in the mid-to-lower portion of the ventricular septum. Pacing parameters (sensing \> 2.0mV; impedance between 300 to 1000Ω; threshold \< 1.5mV) are confirmed to be good.
Beijing Anzhen Hospital, Capital Medical University
Shiyan, Hubei, China
Change in Inter-atrial Electromechanical Delay (Inter AEMD)
Measured by two-dimensional speckle tracking imaging (2D-STI) as the time difference between the onset of the P-wave on the electrocardiogram (ECG) and the peak longitudinal strain of the lateral walls of the right and left atria. A smaller value indicates better inter-atrial synchrony.
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
Change in Intra-atrial Mechanical Delay (IAMD)
Measured by 2D-STI as the standard deviation of the time-to-peak longitudinal strain of six segments of the left atrium. A smaller value indicates better intra-atrial synchrony.
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
Change in Intra-right Atrial Electromechanical Delay (Intra-right AEMD)
Measured by 2D-STI as the time difference between the earliest and latest peak longitudinal strain within the right atrium segments.
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
Change in P-wave Duration (PWD) and Morphology
Measured from the 12-lead surface ECG, including P-wave duration (P-AT, P-Am) and P-wave axis (P-aI, P-aL).
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
Change in QRS Duration and Axis
Measured from the 12-lead surface ECG.
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
Change in Left Atrial (LA) and Right Atrial (RA) Dimensions
Measured by standard 2D echocardiography.
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Change in Left Ventricular End-Systolic Diameter (LVESD) and End-Diastolic Diameter (LVEDD)
Measured by standard 2D echocardiography.
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
Change in Left Atrial Strain Parameters
Includes left atrial reservoir strain (LASr), conduit strain (LAScd), and contraction strain (LASct), measured by 2D-STI.
Time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure
Recurrence of Atrial Arrhythmia
Number of patients experiencing newly developed or recurrent atrial fibrillation or atrial tachycardia.
Time frame: Up to 1 year
Change in Atrial Pacing Threshold
Measured via pacemaker programmer interrogation. Unit: millivolts (mV).
Time frame: 1 Month Post-procedure, 3 Months Post-procedure
Change in Atrial Sensing (P-wave amplitude)
Measured via pacemaker programmer interrogation. Unit: millivolts (mV).
Time frame: 1 Month Post-procedure, 3 Months Post-procedure
Change in Atrial Pacing Impedance
Measured via pacemaker programmer interrogation. Unit: Ohms (Ω).
Time frame: 1 Month Post-procedure, 3 Months Post-procedure
Change in Ventricular Pacing Threshold
Measured via pacemaker programmer interrogation. Unit: millivolts (mV).
Time frame: 1 Month Post-procedure, 3 Months Post-procedure
Change in Ventricular Sensing (R-wave amplitude)
Measured via pacemaker programmer interrogation. Unit: millivolts (mV).
Time frame: 1 Month Post-procedure, 3 Months Post-procedure
Change in Ventricular Pacing Impedance
Measured via pacemaker programmer interrogation. Unit: Ohms (Ω).
Time frame: 1 Month Post-procedure, 3 Months Post-procedure
Percentage of Ventricular Pacing
Measured via pacemaker programmer interrogation. Unit: Percentage (%).
Time frame: 1 Month Post-procedure, 3 Months Post-procedure