To collect data on a variety of market approved Nellcor™ pulse oximetry sensors with the Nellcor™ Pulse Oximetry device to support the market approval.
The purpose of this study is to collect data on the Nellcor™ Pulse Oximetry device when paired with a variety of Nellcor™ Market Released sensors under invasive, controlled desaturation conditions, to verify Pulse Rate and SpO2 accuracy in a diverse participant population over a specified saturation range.
Study Type
OBSERVATIONAL
Enrollment
55
Medtronic Clinical Physiology Lab
Denver, Colorado, United States
Evaluation of pulse rate accuracy
Performance will be determined by calculating the accuracy in terms of root mean square difference (ARMS) comparing the pulse rate measurement from the Nellcor Pulse Oximetry device to the heart rate measurements as obtained by ECG.
Time frame: 2 - 3 hours
Evaluation of saturation accuracy
Performance will be determined by calculating the accuracy in terms of root mean square difference (ARMS) comparing the noninvasive oxygen saturation measurement (SpO2) to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample.
Time frame: 2-3 Months