To validate the NTproBNP2 Test Kit by using the test results of the predicate device as a reference. It will be demonstrated that the investigational product (NTproBNP2 Test Kit) and the predicate are equivalent. Subsequently, it will be demonstrated that the investigational product can be used in clinical testing.
Study Type
OBSERVATIONAL
Enrollment
120
AQT90 FLEX NTproBNP2 Test Kit - Clinical Evaluation - China
Clinical evaluation indicator: The correlation coefficient between the investigational reagent and the predicate. Evaluation criteria: The correlation coefficient between the investigational reagent and the predicate shall not be less than 0.95. The slope should be within the range of \[0.9, 1.1\]. If this requirement is not met, it is recommended to increase the sample size and provide a detailed rationale.
Time frame: 12 months