A single arm study of 20 adult volunteers to assess the safety and efficacy of the ivWatch sensors when observing non-infiltrated (normal) tissues at common peripheral IV therapy sites.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
25
The ivWatch Model 400 monitored tissue at common IV sites over a 24 hour period.
Check IV notification rate per day
The ivWatch Model 400 issues Check IV notifications to communicate the need for a clinician to check an IV site. A red notification suggests an increased likelihood that an IV infiltration may be occurring, with a greater likelihood relative to a yellow notification. This measure describes the average number of red notifications issued by the ivWatch device in a given day when monitoring normal, non-infiltrated tissues. The 95% confidence interval for the red notification rate was calculated using a negative binomial regression model.
Time frame: 24 Hours