The goal of this comparative, cross over single centre study is to to determine whether the average measured vaginal temperature during hyperthermia treatment is (clinically) significant different with the use of a pelotte than with a standard vaginal probe. in 27 women treated with external beam radiotherapy and deep hyperthermie for gynaecological cancer.. The primary endpoint is measured average vaginal temperatures with pelotte versus a standard vaginal probe. The secondary endpoints are: reproducibility of catheter placement, patient comfort, inter-fractional spread of the vaginal temperature and the average vaginal temperature at the tip (0 cm) and 2 cm from the tip with pelotte and a standard vaginal probe. Participants will be treated at least two time with a pelotte and two times with a standard vaginal probe. Participants will be asked to fill in a short questionnaire three times per treatment (before, during and after treatment). A total of 12-15 questionnaires will be completed by the participants over the four to five treatments.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
OTHER
Masking
NONE
Enrollment
27
The pelotte is a medical device. It is a round rod made of PTFE - polytetrafluoroethylene (figure 3). The top 7 cm has a diameter of 20 mm and the bottom 13 cm has a diameter of 11 mm. One disposable catheter for temperature measurements will be attached to the pelotte with tape. This catheter will be placed ventrally on the pelotte. A thermometer will be placed in this catheter. This thermometer will measure the vaginal temperatures. A condom is then placed over the pelotte and the catheter before insertion into the patient. At least two treatment will be with pelotte and at least 2 with a standard vaginal probe.
Erasmus MC
Rotterdam, South Holand, Netherlands
• Average vaginal temperature during DHT treatments with pelotte and with a standard vaginal probe.
To compare if the average vaginal measured temperature is significant different with a pelotte versus the standard vaginal probe
Time frame: approximately seven weeks from the intake with the radiation oncologist to the last treatment administration
Inter-fractional reproducibility of the catheter placement in depth and position with pelotte and with a standard vaginal probe.
For each treatment, the free length of the vaginal probe is measured from the outer labia to the probe tip. This distance is subtracted from 293 mm to determine the internal depth of the vaginal probe. In the control room, the depth is determined based on the temperature measurements. If the measured temperature falls below 36 degrees Celsius at a certain point (e.g., 8 cm from the tip), this point is considered the probe depth. These values are then compared to assess whether there is a difference in reproducibility between treatments using the pelotte and treatments using only a vaginal probe.
Time frame: Approximately seven weeks from the intake with the radiation oncologist to the last treatment administration
Inter-fractional variance in measured vaginal temperatures with pelotte and with a standard vaginal probe.
Time frame: Approximately seven weeks from the intake with the radiation oncologist to the last treatment administration.
Patient-reported comfort with pelotte and with a standard vaginal probe, as assessed using the 'Vragenlijst Hyperthermie Pelotte' questionnaire
Time frame: Approximately seven weeks from the intake with the radiation oncologist to the last treatment administration.
Average vaginal temperature at the measurement points of the tip (0 cm) and two cm from the tip of the catheter with pelotte and with a standard vaginal probe.
Time frame: Approximately seven weeks from the intake with the radiation oncologist to the last treatment administration.