This post-market clinical study was designed to assess the safety and effectiveness of the OpHLINE Ophthalmic Viscosurgical Device (OVD) as an adjunct in cataract surgery. OpHLINE is a sterile, viscoelastic solution based on high-molecular-weight sodium hyaluronate, intended to maintain the anterior chamber space and protect corneal endothelial cells during phacoemulsification and intraocular lens (IOL) implantation. The hypothesis is that OpHLINE, available in three concentrations (1.4%, 2%, and 3%), will provide optimal space maintenance throughout all surgical phases-capsulorhexis, hydrodissection, phacoemulsification, and IOL insertion-while ensuring ease of removal and minimizing postoperative complications. Specifically, the study expects the device to demonstrate high biocompatibility, effective protection of corneal endothelial cells, and a low incidence of adverse events, including intraocular pressure (IOP) spikes. The objectives include confirming compliance with current clinical performance standards and collecting surgeons' feedback on handling and usability. This evaluation aims to strengthen real-world evidence supporting OpHLINE as a safe and reliable solution for cataract surgery, ensuring patient safety and surgical efficiency.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
69
Eligible patients underwent cataract removal by phacoemulsification, followed by implantation of a posterior chamber intraocular lens (IOL) in a single eye. The surgical procedure included the following steps: Capsulorhexis: creation of a continuous circular opening in the anterior capsule of the lens. Hydrodissection: injection of fluid to separate the capsule from the cortex. Phacoemulsification: ultrasonic fragmentation and aspiration of the opacified lens through a micro-incision. IOL insertion: placement of the intraocular lens in the posterior chamber. OpHLINE 1.4% OVD was used as an adjunct during these phases to maintain the anterior chamber space and protect ocular tissues. After lens implantation, the viscoelastic device were aspirated according to standard surgical practice.
Eligible patients underwent cataract removal by phacoemulsification, followed by implantation of a posterior chamber intraocular lens (IOL) in a single eye. The surgical procedure included the following steps: Capsulorhexis: creation of a continuous circular opening in the anterior capsule of the lens. Hydrodissection: injection of fluid to separate the capsule from the cortex. Phacoemulsification: ultrasonic fragmentation and aspiration of the opacified lens through a micro-incision. IOL insertion: placement of the intraocular lens in the posterior chamber. OpHLINE 2% OVD was used as an adjunct during these phases to maintain the anterior chamber space and protect ocular tissues. After lens implantation, the viscoelastic device were aspirated according to standard surgical practice.
Eligible patients underwent cataract removal by phacoemulsification, followed by implantation of a posterior chamber intraocular lens (IOL) in a single eye. The surgical procedure included the following steps: Capsulorhexis: creation of a continuous circular opening in the anterior capsule of the lens. Hydrodissection: injection of fluid to separate the capsule from the cortex. Phacoemulsification: ultrasonic fragmentation and aspiration of the opacified lens through a micro-incision. IOL insertion: placement of the intraocular lens in the posterior chamber. OpHLINE 3% OVD was used as an adjunct during these phases to maintain the anterior chamber space and protect ocular tissues. After lens implantation, the viscoelastic device were aspirated according to standard surgical practice.
Basurto University Hospital
Bilbao, Bizkaia, Spain
Araba University Hospital
Vitoria-Gasteiz, Álava, Spain
Ability to maintain anterior chamber space during cataract surgery.
This outcome evaluates whether the OpHLINE OVD can maintain a stable and adequate anterior chamber throughout all critical phases of cataract surgery: capsulorhexis, hydrodissection, phacoemulsification, and intraocular lens (IOL) insertion. Surgeons assessed this parameter using a structured questionnaire, classifying the chamber status into four categories: Complete chamber, Adequate working space, Shallow chamber, and Flat. A positive outcome is defined as achieving either a complete chamber or adequate working space in at least 70% of cases. This measure reflects the core functionality of the OVD in providing surgical safety and facilitating precise maneuvers.
Time frame: 90 days
Safety evaluation of OpHLINE during and after surgery
This measure involves systematic monitoring and documentation of any serious adverse events related to the device during the entire study period. SAEs include any event that could result in significant harm or require medical intervention. The target is an incidence rate of ≤1%, confirming that the device does not introduce unexpected risks beyond those inherent to cataract surgery.
Time frame: 90 days
Assessment of removal difficulty after IOL implantation
Surgeons classify removal as Easy, Normal, or Problematic. Adequate removal is defined as Easy or Normal in at least 70% of cases
Time frame: 6 hours
Percentage of endothelial cell density loss
Endothelial cell density was measured using a non-contact specular microscope at baseline and at 90 days post-surgery. The goal is to determine the protective effect of the OVD on corneal tissue during surgery. A loss of ≤12% is considered optimal, as excessive cell loss can compromise corneal transparency and visual outcomes.
Time frame: 90 days
Incidence of IOP ≥30 mmHg during follow-up
IOP was measured at all scheduled visits to monitor postoperative pressure changes. The target is ≤13% incidence at 6 hours post-surgery and ≤6% at subsequent visits. Elevated IOP (\>30 mmHg) can lead to optic nerve damage if not controlled, so this measure ensures the device does not contribute to clinically significant pressure spikes.
Time frame: 90 days
Frequency of non-serious adverse events during follow-up
This includes any minor complications or events that occur during or after surgery, such as mild inflammation or transient discomfort. The acceptable threshold is ≤30% across the study period, aligning with typical postoperative profiles for cataract surgery.
Time frame: 90 days