Researchers at the Department of Intensive Internal Medicine and the Department of Endocrinology and Diabetology will conduct a study to verify the accuracy and safety of the new Syai Tag system for continuous glucose monitoring in patients in intensive internal care. Almost 150,000 people in Slovenia have diabetes, so keeping a close eye on blood sugar levels is key to preventing complications. The new sensors for continuous monitoring are available over the counter and certified as medical devices. The study will include at least 100 patients who will need blood sugar monitoring during their stay in the intensive care unit. Each patient will have two small sensors placed on each upper arm to continuously measure their blood sugar levels, but these values will not be visible to medical staff. At the same time, healthcare professionals will perform routine blood sugar measurements, and researchers will then compare the accuracy of both methods. The procedure is safe and painless, and patients will receive the same quality of care as usual.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DEVICE_FEASIBILITY
Masking
NONE
Enrollment
100
patients who will be hospitalized in the intensive internal medicine department and will receive CGM for glucose monitoring
UMC Maribor
Maribor, Slovenia
RECRUITINGPercentage of sensor functional uptime during the ICU stay.
The proportion of time during which the CGM system continuously provided glucose data without signal loss or mechanical sensor failure.
Time frame: From enrollment to discharge or up to 14 days.
Mean Absolute Relative Difference (MARD) between CGM and arterial blood gas glucose measurements.
Statistical comparison between CGM values and reference laboratory measurements from arterial blood to assess the accuracy of the device.
Time frame: From admission to discharge or up to 14 days.
Impact of noradrenaline on Mean Absolute Relative Difference (MARD) between CGM and arterial glucose measurements.
The MARD will be calculated to assess CGM accuracy compared to the gold standard (arterial blood gas). To evaluate the impact of noradrenaline, MARD will be stratified and compared between two patient groups: those receiving noradrenaline (Yes) and those not receiving noradrenaline (No). Unit of Measure: Percentage (%).
Time frame: From enrollment to discharge or up to 14 days.
CGM lifespan in ICU conditions.
The effective duration of CGM sensor function will be evaluated and compared to the manufacturer's expected lifespan.
Time frame: From enrollment to discharge or up to 14 days.
CGM glucose measurements are comparable to arterial blood glucose measurements.
We will measure glucose in interstitial fluid with a CGM, expressed in mmol/l, and obtain arterial blood to determine blood glucose in mmol/l, and compare the analyses.
Time frame: From enrollment to discharge or up to 14 days.