The goal of this study is to determine if the Lactate Plus meter is accurate compared to lab lactate levels, and to determine if the Accu chek guide glucometer is accurate compared to lab serum glucose levels in patients with Glycogen Storage Disease Types Ia, Ib and XI. To determine this, patient's will have a one-time planned admission to Connecticut Children's for approximately 8 hours and receive hourly blood draws as well as finger-sticks.
Study Type
OBSERVATIONAL
Enrollment
10
Hourly blood lactate levels
Hourly blood glucose levels
Connecticut Children's
Hartford, Connecticut, United States
RECRUITINGTo determine if the lactate Plus meter (both capillary and serum sample) readings are within 20% of lab serum lactate reading 95% of the time, in a population of patients with glycogen storage disease type Ia/Ib/XI
Time frame: Hourly x 8 hours
To determine if the Accu-Chek Guide glucometer (capillary sample) readings are within 20% of lab serum glucose reading 95% of the time, in a population of patients with glycogen storage disease type Ia/Ib/XI
Time frame: Hourly x 8 hours