This is a prospective non-randomized patient controlled single center study which is designed to evaluate the clinical outcomes of the treatment of an implanted intraocular lens using a low-energy femtosecond laser for refractive correction.
The purpose of this research study is to evaluate the safety and efficacy of the treatment of an implanted intraocular lens with a device. The device employs a generic femtosecond laser used in numerous other ophthalmic procedures. The device uses low levels of energy from the femtosecond laser to correct spherical, cylindrical and sphero-cylindrical refractive error existing in an implanted intraocular lens. The refractive correction is effected through the treatment, which is employed at least 30 days after cataract surgery.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
12
The Perfector is used in the procedure. The device is attached to the patient by the use of a patient attachment. The patient attachment attaches to the sclera of the eye using vacuum pressure.
Panama Eye Center, S.A. (formerly known as Clinica de Ojos Orillac - Calvo)
Panama City, Panama
Primary Endpoint - Change in mean manifest refraction spherical equivalent ("MRSE") (post- versus pre-treatment)
Change in mean manifest refraction spherical equivalent ("MRSE") post-treatment at 7-day visit versus pre-treatment.
Time frame: Between pre-treatment and 1 week post-treatment.
Safety Endpoint - Primary Device Related Adverse Events (post- versus pre-treatment)
Demonstrate that mesopic contrast sensitivity (with and without glare) post-treatment (at 7, 30 and 90-day follow up visits) is non-inferior to pre-treatment as measured by CTS contrast sensitivity (NI margin of 0.15 log units).
Time frame: Between pre-treatment and 1 week, 1 month and 3 months post-treatment.
Safety Endpoint - Primary Device Related Adverse Events (post- versus pre-treatment)
Color vision discrimination compared between pre-op and post-op at 90-day visit (measured by the FM 100 Hue test) comparison of visual disturbances between pre-treatment and post-treatment (at 30 and 90 day-day follow up) assessed via Quality of Vision (QoV) questionnaire.
Time frame: Between pre-treatment and 1 month and 3 months post-treatment.
Secondary Endpoint - Demonstrate Best Corrected Distance Visual Acuity (BCDVA) (post- versus pre-treatment)
Demonstrate that Best Corrected Distance Visual Acuity (BCDVA) at 4m post-treatment (at 90-day visit) is non-inferior to pre-treatment BCDVA (NI margin of 0.1 logMAR, SD=0.12).
Time frame: Between pre-treatment and 3 months post-treatment.
Secondary Endpoint - Comparison of manifest refraction spherical equivalent (MRSE) (post- versus pre-treatment)
Comparison of manifest refraction spherical equivalent (MRSE) post-treatment at 7 days, 30 days, and 90 days versus pre-treatment intended MRSE.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: Between pre-treatment and 1 week, 1 month and 3 months post-treatment.
Secondary Endpoint - Manifest refraction spherical equivalent (MRSE) stability (between any two follow up visits)
Manifest refraction spherical equivalent (MRSE) remains stable (i.e. within +/-0.5 diopter between any two follow up study visits).
Time frame: Between pre-treatment and 1 week, 1 month and 3 months post-treatment.