This study aimed to evaluate the feasibility and safety, and to observe early efficacy signals of the focused ultrasound mediated drug delivery system combined with SonoVue® in patients with HER2-positive breast cancer receiving neoadjuvant therapy by comparing its pathological complete response (pCR) rate with a matched historical cohort of HER2-positive breast cancer patients treated at our center.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
20
Participants will undergo six cycles of focused ultrasound-mediated targeted drug delivery combined with neoadjuvant therapy. The ultrasound parameters are set as follows: ultrasound power 9 W (acoustic pressure 3 MPa), frequency 1 MHz, pulse duration 100 μs, pulse repetition frequency 100 Hz, with 5 treatment cycles per single point. During each treatment session, 2.5 mL of SonoVue® ultrasound contrast agent microbubbles is injected via the cubital vein, followed by a 5 mL normal saline flush. Subsequently, 2.5 mL of SonoVue® microbubbles are administered every 2 minutes, with a total injection volume of 15 mL throughout the procedure.
The Second Affiliated Hospital Zhejiang University School of Medicine (SAHZU)
Hangzhou, Zhejiang, China
RECRUITINGPathological Complete Response rate
Time frame: 18 weeks from the enrollment, after the surgey
Tumor Pathological Complete Response
Time frame: 18 weeks from the enllorment, after the surgey
Event-Free Survival (EFS)
Time frame: 3 years after the surgery