This multicenter, prospective, randomized controlled trial aims to evaluate whether a novel multifunctional oropharyngeal airway (MOPA) reduces the incidence of hypoxemia in 1,518 adult patients (ASA I-II, aged 18-80 years) undergoing elective sedated gastrointestinal endoscopy. Patients are randomized 1:1 to receive either the MOPA (which integrates oxygen delivery, PETCO₂ monitoring, and airway support) or conventional nasal cannula with standard mouthpiece. The primary endpoint is the incidence of hypoxemia (75% ≤ SpO₂ \< 90% for \<60 seconds) during the procedure. Secondary outcomes include severe hypoxemia, hypercapnia, PETCO₂ monitoring success, airway interventions, adverse events, and satisfaction scores. The study is conducted across 29 centers in China, with centralized randomization via an EDC system, blinded outcome assessment, and statistical analysis using a two-sided alpha of 0.05 (power 90%). Results are expected to provide high-level evidence for optimizing airway management during sedated endoscopy.
Study Title Effect of a Novel Multifunctional Oropharyngeal Airway on Hypoxemia in Patients Undergoing Sedated Gastrointestinal Endoscopy: A Multicenter, Prospective, Randomized Controlled Trial Principal Investigator Dr. Wu Jianbo, Department of Anesthesiology and Perioperative Medicine, The First Affiliated Hospital of Shandong First Medical University (Shandong Provincial Qianfoshan Hospital) Participating Centers A total of 29 centers across China (including Zhejiang University First Affiliated Hospital, Hebei Medical University Second Hospital, Qilu Hospital of Shandong University, etc.) Study Objective To evaluate whether a novel multifunctional oropharyngeal airway (MOPA) reduces the incidence of hypoxemia compared to conventional nasal cannula with standard mouthpiece in patients undergoing sedated gastrointestinal endoscopy. Study Design Multicenter, prospective, parallel-group, randomized controlled trial. Sample Size 1,518 patients (759 per group), accounting for a 20% dropout rate. Power: 90%, two-sided alpha: 0.05. Participant Criteria Inclusion: Age 18-80 years; ASA I-II; scheduled for elective sedated gastroscopy + colonoscopy; written informed consent. Exclusion: Known respiratory disease (asthma, COPD, moderate-severe OSA, etc.); bleeding tendency or oral/nasal mucosal injury; severe cardiac/renal/hepatic dysfunction; therapeutic endoscopy (e.g., polypectomy, EMR); pregnancy; known drug allergy; emergency procedure; alcohol abuse; psychiatric illness; myasthenia gravis; participation in another trial within 3 months; refusal. Interventions Experimental group (MOPA): After standard sedation (fentanyl 0.05 μg/kg + propofol 1.5-2.5 mg/kg, then propofol infusion 4-12 mg/kg/h), a novel multifunctional oropharyngeal airway (integrating a bite block, oropharyngeal airway, oxygen delivery channel, and CO₂ sampling port) is inserted. Oxygen 3-4 L/min is delivered via the device, and PETCO₂ is continuously monitored through the sampling port. Control group (conventional mouthpiece): Same sedation regimen. Patients receive oxygen 3-4 L/min via nasal cannula. PETCO₂ is monitored via nasal cannula sampling line. No oropharyngeal airway is used. Both groups receive standardized rescue interventions for hypoxemia or ventilatory abnormalities. Primary Endpoint Incidence of hypoxemia defined as 75% ≤ SpO₂ \< 90% lasting \< 60 seconds during the procedure. Secondary Endpoints Severe hypoxemia (SpO₂ \< 75% OR 75% ≤ SpO₂ \< 90% lasting ≥ 60 s) PETCO₂ monitoring success rate (proportion with clear continuous waveform) Detection rate of hypoventilation events (apnea ≥15-20 s; respiratory rate \<8/min for ≥30 s) and early warning time before SpO₂ decline Incidence of hypercapnia (PETCO₂ \>50 mmHg or \>45 mmHg for ≥30 s) Adverse events: airway injury, regurgitation, aspiration, coughing, laryngospasm, post-procedural sore throat, hoarseness, dental/oral soft tissue injury Airway intervention burden (jaw thrust, increased O₂ flow, procedure pause, face-mask ventilation, laryngeal mask/intubation) Procedural efficiency (scope insertion-to-withdrawal time, number of interruptions/withdrawals due to respiratory events) Endoscopist satisfaction (0-10 scale) and patient satisfaction (customized 0-2 per item score) Safety Outcomes Device-related (mucosal injury, bleeding, sore throat), respiratory events (cough, laryngospasm, aspiration, need for advanced airway), systemic complications (hemodynamic instability, unplanned hospitalization/ICU), serious adverse events (refractory hypoxemia, emergency airway support, arrhythmia, myocardial ischemia, anaphylaxis, stroke). Statistical Analysis Primary endpoint comparison using chi-square or Fisher's exact test; continuous variables using t-test or Mann-Whitney U test as appropriate. All tests two-sided, significance level α=0.05. Analysis sets: FAS, PP, SS. Missing data handled by multiple imputation where applicable. Randomization and Blinding Centralized randomization via EDC system using stratified block randomization (block sizes 4,6,8; stratified by center). Patients are blinded to group assignment. Operators are unblinded due to device appearance. Outcome assessors and statisticians are blinded until database lock (two-stage unblinding). Study Duration Start anticipated June 2026, completion December 2027. Expected Publications 1-2 SCI-indexed papers reporting the primary and secondary outcomes.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
PREVENTION
Masking
TRIPLE
Enrollment
1,518
The Novel Multifunctional Oropharyngeal Airway is an integrated airway device designed for sedated gastrointestinal endoscopy. It combines a modified mouthpiece, an oropharyngeal airway, an oxygen delivery channel, and a PETCO₂ sampling port into a single unit. The device maintains upper airway patency by preventing tongue prolapse, delivers oxygen directly to the pharynx near the glottis for more efficient oxygenation, and enables continuous real-time capnography monitoring. It is made of soft medical-grade material with anatomical curvature and depth markings to minimize mucosal injury and patient discomfort. The MOPA also reserves an interface for emergency jet ventilation if needed. It is inserted under sedation and allows the endoscope to pass smoothly through its central channel without obstruction.
Incidence of intraoperative hypoxemia (75% ≤ SpO₂ < 90%, duration < 60 seconds)
Incidence of intraoperative hypoxemia (75% ≤ SpO₂ \< 90%, duration \< 60 seconds)
Time frame: Perioperative
Incidence of severe hypoxemia (SpO₂ < 75% OR 75% ≤ SpO₂ < 90% lasting ≥ 60 seconds).
Incidence of severe hypoxemia (SpO₂ \< 75% OR 75% ≤ SpO₂ \< 90% lasting ≥ 60 seconds).
Time frame: Perioperative
PETCO₂ monitoring success rate
proportion of participants from whom a clear, continuous PETCO₂ waveform is obtained
Time frame: Perioperative
Detection rate of hypoventilation events
apnea (waveform disappearance ≥ 15-20 seconds) / bradypnea (respiratory rate \< 8 breaths/min lasting ≥ 30 seconds), etc.; and the proportion of early recognition before SpO₂ decline, as well as the advanced warning time (data exported from the monitor)
Time frame: Perioperative
Incidence of hypercapnia
PETCO₂ \> 50 mmHg (or \> 45 mmHg) lasting ≥ 30 seconds
Time frame: Perioperative
Incidence of adverse events
Record all adverse events that occur during the experiment for final analysis
Time frame: Perioperative
Airway intervention burden
jaw thrust/chin lift
Time frame: Perioperative
Procedural efficiency
Number of withdrawals due to interruption caused by respiratory events
Time frame: perioperative
Satisfaction: endoscopist satisfaction and patient satisfaction.
Note - Patient satisfaction: Each participant, in the absence of research personnel, will provide quantitative scores on predefined dimensions (pain level, foreign body sensation, comfort, quality of recovery, and overall experience). Each item is scored from 0 to 2 points, and the sum of the scores on all items will be used as the final satisfaction assessment. Endoscopist satisfaction: This is assessed using a 0-10 point scale, where 0 represents "very dissatisfied" and 10 represents "very satisfied". Based on their actual experience during the procedure, the endoscopist will select the corresponding number on the rating scale to indicate their satisfaction with the participant's cooperation and response throughout the entire procedure.