The purpose of this study is to collect NIBP and invasive aortic pressure data from the clinical setting.
The purpose of the study is to collect clinical data for validation of the accuracy of non-invasive blood pressure (NIBP) measurement algorithms on the Non-invasive Blood Pressure (NIBP) Standard Parameter Module (SPM) per IEC 80601-2-30:2018, Medical electrical equipment-Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers standard, section 201.106. In addition, the study will be conducted per ISO 81060-2:2018+AMD2:2024 Non-invasive sphygmomanometers-Part 2: Clinical investigation of intermittent automated measurement type.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
15
The collection of non-invasive blood pressure measurements using NIBP Standard Parameter Module with SuperSTAT algorithm with simultaneous invasive blood pressure waveform recordings per discretion of the sponsor upon combining the data from other clinical studies to comply with ISO 81060-2:2018 +AMD2:2024.
Tampere Heart Hospital
Tampere, Finland
RECRUITINGPrimary Outcome Measure
Collection of non-invasive blood pressure measurements data from subjects in clinical setting.
Time frame: Up to 2 hours after time of enrollment