The purpose of this clinical study is to evaluate the effectiveness and safety of a specialized medical device, the Naviscore scoring balloon, in preparing calcified coronary artery narrowings before the implantation of a drug-eluting stent. During percutaneous coronary interventions, the presence of calcified plaques in the heart arteries represents a major challenge because it can prevent stents from expanding fully. When a stent remains under-expanded, it significantly increases the long-term risk of arterial re-narrowing or blood clot formation. To optimize stent expansion, appropriate preparation of the diseased vessel section before stent insertion is a critical phase. This study is a prospective, multi-center randomized trial designed to test the hypothesis that treating calcified coronary lesions with the Naviscore scoring balloon will achieve a better stent expansion and a larger final minimal stent area compared to standard lesion preparation using regular non-compliant balloons. Eligible participants will be randomized in a one-to-one ratio to one of these two lesion preparation strategies. For all included patients, standard drug-eluting stents will be deployed. The study will use intravascular ultrasound imaging to evaluate the final minimum area of the stent directly inside the treated artery at the site of the highest initial calcium burden. Participant health and clinical outcomes will be monitored for up to twelve months following the procedure.
Percutaneous transluminal coronary angioplasty (PTCA) is the main technique in percutaneous coronary intervention (PCI). It aims to restore coronary blood flow by mechanically enlarging a stenotic arterial segment through controlled balloon inflation within the vessel lumen. The acute luminal gain with balloon angioplasty is the result of multiple mechanisms, ranging from percolation and longitudinal redistribution of atheroma, to plaque fracture, and overstretching of the vessel wall, often accompanied by intimal tearing and medial dissection. While contemporary drug-eluting technologies have significantly improved long-term clinical outcomes, fibrocalcified atheromatous intracoronary plaques still present a significant challenge for stent crossing and optimal deployment. Research has demonstrated that suboptimal stent expansion and malapposition are significant contributing factors to restenosis and stent thrombosis rates, regardless of the type of stent used. Intravascular imaging registries have identified that twenty to thirty percent of deployed stents remain under-expanded or malapposed in daily clinical practice. Severe coronary calcification can make it difficult to deliver devices and can limit the expansion of stents, which can result in suboptimal improvement in blood flow and an increased risk of adverse events, including target lesion failure. Furthermore, heavily calcified lesions have been found to be associated with higher rates of vessel dissection, perforation and impaired anti-proliferative drug delivery, resulting from the mechanical barrier of the calcium burden. In order to achieve successful stent deployment, optimal preparation of fibro-calcific lesions is essential and involves a number of plaque modification strategies. Focused force angioplasty is a procedure that targets specific locations of the plaque, focusing inflation forces on the area to induce localized stress and facilitate cracking. It is crucial to ensure that the pre-dilatation balloon is correctly sized to maximise safety by minimising the risk of advanced coronary perforation or severe medial damage. It is imperative that sizing compliance chart data is followed rigorously, as compliant balloons can display an unpredictable non-linear expansion behaviour at higher pressures, whereas non-compliant balloons maintain a stable volume, concentrating their dilating force directly at the calcified site up to their rated burst pressure. Semi-compliant pre-dilatation balloons often expand asymmetrically in resistant lesions, increasing the risk of edge dissections. The sizing and performance of scoring devices and non-compliant balloons remains limited in contemporary clinical literature. This randomised trial is designed to evaluate the post-market clinical performance, safety and comparative effectiveness of the Naviscore scoring balloon catheter manufactured by iVascular.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
200
Percutaneous coronary preparation of calcified coronary stenosis using the rapid exchange Naviscore scoring balloon catheter before drug-eluting stent deployment. The device incorporates an over-the-balloon metallic nitinol element designed to focally concentrate dilation forces, minimizing balloon slippage and scoring the atheromatous plaque at lower pressures to facilitate optimal stent expansion. Sizing is based on a 0.8:1 to 1:1 ratio relative to the reference lumen. Operators must execute a specific sequence of inflating the scoring balloon to nominal pressure, fully deflating it, mobilizing it within the lesion, and reinflating it, repeated three to four times across the target segment to ensure multi-focal plaque disruption before stenting.
Conventional non-compliant (NC) PTCA balloon catheter used for standard lesion pre-dilatation according to routine clinical practice, inflated below the rated burst pressure without parallel guidewires in place.
CHU - HELORA site Kennedy
Mons, Hainaut, Belgium
Imelda
Bonheiden, Belgium
Hôpital Erasme - Cliniques universitaires de Bruxelles
Brussels, Belgium
Hôpital Vivalia
Libramont, Belgium
CHR de la Citadelle
Liège, Belgium
Humani CHU Charleroi - Chimay
Lodelinsart, Belgium
INCCI Haerz Zenter
Luxembourg, Luxembourg
Centro Hospitalar de Vila Nova de Gaia/Espinho
Vila Nova de Gaia, Portugal
Hospital de Sant Pau
Barcelona, Spain
Hospital Universitari de Bellvitge
Barcelona, Spain
...and 4 more locations
Final Minimal Stent Area (MSA)
The minimal stent area (MSA) will be measured in square millimeters by intravascular ultrasound (IVUS) after final procedural optimization. The primary comparison will evaluate the MSA specifically at the exact site of the initial heaviest calcium burden to demonstrate the superiority of the Naviscore scoring balloon over standard non-compliant (NC) balloons.
Time frame: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure.
Minimal Stent Area (MSA) Across the Entire Stented Segment
The minimal stent area (MSA) in square millimeters measured across the entire length of the deployed segment using intravascular ultrasound (IVUS) pullbacks to evaluate overall stent expansion expansion
Time frame: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent Eccentricity at the Site of the Minimal Stent Area
Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) at the exact cross-section of the minimal stent area segment
Time frame: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent Eccentricity at the Initial Heaviest Calcium Burden Site
Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) specifically at the cross-section matching the initial heaviest baseline calcium plaque burden
Time frame: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Device Success Rate
Percentage of cases achieving successful delivery of the assigned pre-dilatation balloon, complete inflation with adequate lesion preparation, and the absolute absence of vessel rupture or advanced coronary perforation categorized by the Ellis classification
Time frame: During the index percutaneous coronary intervention (PCI) procedure
Major Adverse Cardiac Events (MACE) Rate
Percentage of participants experiencing Major Adverse Cardiac Events (MACE), defined as a composite clinical endpoint of myocardial infarction, stroke or transient ischemic attack, or cardiac death
Time frame: At 1 and 12 months post-procedure
Target Lesion Failure (TLF)
Percentage of participants experiencing Target Lesion Failure (TLF), defined as a composite of cardiac death, target-vessel myocardial infarction defined according to ARC-2 criteria, or clinically indicated Target Lesion Revascularization
Time frame: At 1 and 12 months post-procedure
Target Vessel Failure (TVF)
Defined as a composite of cardiac death, target-vessel myocardial infarction, or clinically indicated Target Vessel Revascularization (TVR).
Time frame: At 1 and 12 months post-procedure
Clinically Indicated Target Lesion Revascularization (TLR)
Percentage of participants requiring a repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) driven by a recurrent narrowing inside the originally treated target lesion segment (including the stent and its five-millimeter proximal and distal borders), which is accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.
Time frame: At 1 and 12 months post-procedure
Clinically Indicated Target Vessel Revascularization (TVR)
Percentage of participants requiring any repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) in any segment of the originally treated target vessel, driven by angiographic restenosis or disease progression, and accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.
Time frame: At 1 and 12 months post-procedure
Procedural Success Rate
Percentage of cases achieving Device Success (successful delivery of the assigned balloon and complete inflation with adequate lesion preparation) combined with a final residual stenosis of less than thirty percent as measured by quantitative coronary angiography (QCA) and a final Thrombolysis in Myocardial Infarction (TIMI) grade 3 flow, without the occurrence of in-hospital Major Adverse Cardiac Events (MACE).
Time frame: During the index percutaneous coronary intervention (PCI) procedure and until hospital discharge (up to 24 hours post-procedure).
Final Residual Stent Underexpansion
Evaluation of the final residual underexpansion of the drug-eluting stent after deployment and high-pressure post-dilatation, measured in millimeters using motion-corrected X-ray stent visualization software (StentBoost or equivalent stent enhancement method) in two orthogonal views without contrast.
Time frame: At the end of the index percutaneous coronary intervention (PCI) procedure (immediate post-stenting optimization).
Final Stent Diameter Stenosis
The final percentage of stent diameter stenosis evaluated by quantitative coronary angiography (QCA) at the index procedure after final stent deployment and post-dilatation optimization.
Time frame: At the end of the index percutaneous coronary intervention (PCI) procedure.
Rate of Procedural Complications
The cumulative rate of procedure-related complications during the index procedure, defined as a composite of cardiac death, acute myocardial infarction, emergency coronary artery bypass graft surgery (CABG), coronary artery perforation (Ellis classification), and cardiac tamponade directly related to the use of the Naviscore scoring balloon or the regular non-compliant (NC) balloon.
Time frame: During the index percutaneous coronary intervention (PCI) procedure (immediate intra-procedural period).
Procedural Cost-Effectiveness
Evaluation of the procedural cost-effectiveness of using the Naviscore scoring balloon compared to standard non-compliant (NC) balloons for calcified lesion preparation, calculated by factoring device costs, overall material utilisation during the intervention, procedure duration, and the need for bailout additional plaque-modification tools or complementary high-pressure devices.
Time frame: During the index percutaneous coronary intervention (PCI) procedure.