This study will evaluate whether gentle skin abrasion improves electromyographic signal quality during neuromuscular blockade monitoring in adult surgical patients. Participants will receive standard TwitchView electromyographic neuromuscular blockade monitoring on one wrist and TwitchView monitoring after gentle skin abrasion on the opposite wrist. The study will compare electromyographic response amplitudes and supramaximal stimulus currents between routine electrode application and electrode application after skin preparation. The study procedures will occur during scheduled surgery and will not alter the anesthetic or surgical management of participants.
Neuromuscular blocking agents are commonly used during adult surgical procedures. Quantitative neuromuscular blockade monitoring is used to assess the effects of these drugs and to guide clinical management of neuromuscular blockade and recovery. The TwitchView electromyographic monitor is routinely used for this purpose and uses surface electrodes placed over the ulnar nerve and hand muscles to generate train-of-four measurements. In some patients, the electromyographic signal obtained with routine electrode application may be poor or inadequate for continuous monitoring. This may be related to skin condition, such as callouses, oily skin, or other factors that affect signal quality. Gentle skin abrasion is a technique commonly used in electrophysiology settings to improve the quality of physiologic electrical signals. This study will compare electromyographic signals obtained from routinely applied TwitchView electrodes with signals obtained from TwitchView electrodes placed after gentle skin abrasion using 3M Red Dot Trace Prep abrasive strips. After consent is obtained and the participant is moved into the operating room, TwitchView electrodes will be applied to both wrists. One wrist will receive routine electrode application without skin preparation, while the opposite wrist will undergo gentle skin abrasion before standard TwitchView electrodes are applied. Two TwitchView monitors will be used, one for each electrode strip. After induction of anesthesia, both TwitchView devices will be started. Other than a brief device start-up period, anesthetic and surgical management, including neuromuscular blocking drug dosing and timing, will be determined by the clinical care team and not by the investigators. The devices will record neuromuscular monitoring data during the procedure. The primary outcome is the amplitude of recorded electromyographic responses to train-of-four stimulation after induction of anesthesia but before administration of a paralytic drug. Secondary outcomes include the supramaximal stimulus current determined by the device and electromyographic signal amplitudes after administration of neuromuscular blocking drugs.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
55
Gentle skin abrasion will be performed at the stimulus and recording electrode sites on one wrist using 3M Red Dot Trace Prep abrasive strips before applying standard TwitchView electrodes. The opposite wrist will receive routine electrode application without skin preparation. Two TwitchView monitors will be used to compare EMG signal amplitudes and supramaximal stimulus currents.
University of Minnesota East Bank Medical Center
Minneapolis, Minnesota, United States
EMG Response Amplitude Before Neuromuscular Blocking Drug Administration
Amplitude of recorded EMG responses to train-of-four stimulation will be compared between routine electrode application and electrodes placed after gentle skin abrasion.
Time frame: During surgery, before paralytic drug administration
Supramaximal Stimulus Current
Supramaximal stimulus current determined by the TwitchView device will be compared between routine electrode application and electrodes placed after gentle skin abrasion.
Time frame: During device startup after anesthesia induction
EMG Response Amplitude After Neuromuscular Blocking Drug Administration
EMG signal amplitudes after administration of neuromuscular blocking drugs will be compared between routine electrode application and electrodes placed after gentle skin abrasion.
Time frame: During surgery, after paralytic drug administration